Zeta Potential Analyzers, classification and areas of
applicability.
Zeta potential analyzers is an analytical device for
characterizing zeta potential as a property of interfacial
layer in liquid particulates (dispersions, emulsions) and
porous bodies. There is
ISO Standard -13099
being prepared that would describe the
main principles underlying functioning of such devices.
Zeta potential
is not directly measurable parameter.
It should be calculated from other experimentally measurable
properties of the system using appropriate theory. There is
no Zeta Potential analyzer that would be capable of working
with all possible liquid based heterogeneous systems. All of
them have limitation. It might be quite challenging to find
an optimum device for particular application. Here we
present some brief classification of these devices pointing
out some obvious restrictions.
All these devices must have means for moving liquid
relative to the solid surface for disturbing double layer
(1) and also
means for monitoring generated signals
(2).
(1). Relative motion of the
liquid can be induced by external electric field
or mechanical pressure field.
Instruments relying on electric field for
this purpose are based on phenomenon of Electrophoresis
(dilute systems) or ElectricSonic Amplitude
(concentrated systems). Electrophoresis based instruments
apply DC or low frequency (<10 KHz) AC field. In contrary,
ESA instruments apply high frequency AC field on scale of
MHz, usually below 10 MHz. Practically all such
devices are suitable for liquid particulate systems, not for
porous bodies.
Instrument relying on mechanical pressure field
are suitable for both, particulate systems and porous
bodies. In the case of particulates a high frequency
ultrasound (MHz range) is used for moving particles
relative to the liquid. This is so-called Colloid
Vibration Current effect. In the case of porous body
constant gradient of pressure could move liquid relative to
the porous matrix, which generates Streaming current
in isochoric mode. Propagation of ultrasound through
a porous body generates electric signal as well, known as Seismoelectric effect. It can be defined as
non-isochoric streaming current. These instruments can
be used only with concentrated systems. There are several
USA Patents (see below [1-3]) describing such instruments.
There is an approximate threshold between diluted and
concentrated system : 1 % vl.
(2). Monitoring can be
achieved using optical means, electrical
or mechanical.
Optical devices use either dynamic
light scattering (ELS) or image analysis
for extracting information on the speed of the particles
motion in the electric field. These instruments are suitable
only for dilute particulate systems. Instrument design
should prevent liquid from thermal motion, which usually
restricts high limit of the ionic strength. Possible
electrosmotic flow on the measuring chamber walls is another
complicated factor, especially for image analysis devices.
Electric output is used in streaming current/potential
devices for porous bodies and in the CVI Zeta probe for both,
particulates and porous bodies. Streaming current is
monitored usually in DC mode or at very low frequency. CVI,
on the other side is high frequency AC current, which
requires measurement of the magnitude and phase.
Mechanical output is used in form of
measured ultrasound signal generated by ESA devices.
References:
-
Dukhin, A.S. and Goetz, P.J. “Method and device for characterizing
particle size distribution and
zeta potential in concentrated system by
means of Acoustic and Electroacoustic Spectroscopy”, patent USA,
6,109,098 (2000)
-
Dukhin, A.S. and Goetz, P.J. “Method and device for Determining
particle size Distribution and
zeta potential in Concentrated
Dispersions”, patent USA, 6,449,563 (2002)
-